Audiology and Speech Research > Volume 12(1); 2016 > Article
Audiology and Speech Research 2016;12(1): 12-23.
Published online: January 31, 2016.
doi: http://doi.org/10.21848/asr.2016.12.1.12
순음청력검사기의 교정
김진동
부산가톨릭대학교 보건과학대학 언어청각치료학과
Audiometric Calibration of Pure Tone Audiometers
Jin-Dong Kim
Department of Speech and Hearing Therapy, College of Health Science, Catholic University of Pusan, Busan, Korea
Correspondence  Jin-Dong Kim ,Tel: +82-51-510-0844, Fax: +82-51-510-0848, Email: jdkim@cup.ac.kr
Received: November 12, 2015; Revised: December 28, 2015   Accepted: January 3, 2016.  Published online: January 31, 2016.
ABSTRACT
Purpose:
This review is to introduce the performance characteristics of pure tone audiometer specified in the standards and necessary basic calibration instrumentations, explain the electroacoustic calibration among calibration procedures in order to determine if the pure tone audiometer which is being used currently satisfy the standards.
Method:
Audiometric calibration consists of three parts; measurement of audiologic equipment performance characteristics, comparison of the measurements to a standard specified in the international or national standards, and adjustment as needed. Calibration checks and adjustments are an essential process for reducing a measurement errors between audiologic equipment and improving the reliability and accuracy of audiologic test result.
Results:
Specific calibration items that should be checked from audiologic equipment are summarized in International Electrotechnical Commission, International Organization for Standardization, and American National Standards Insti-tute standards. Audiologists should use audiologic equipment which meets appropriate standards and carry out the calibration in order to determine whether the result of calibration check satisfies current standards and exists in the range of a specific tolerances. Reliable test results obtained with audiology equipments calibrated support best audiology practice leading to increased confidence of audiologists.
Conclusion:
Audiologists have a responsibility to either check its calibration personally or to arrange for periodic calibration of the equipment by an outside service in order to maintain the accuracy of audiologic equipment. For this, audiologists should have proper ability to check whether audiologic equipment meets to the standardized performance characteristics. However, the performance characteristics and calibration methods vary according to the type of audiologic equipment.
Key Words: Audiometric calibration, Standards, Pure tone audiometer, Electroacoustic calibration.
중심단어: 청력검사기의 교정· 표준· 순음청력검사기· 전기음향교정.
TOOLS
PDF Links  PDF Links
Full text via DOI  Full text via DOI
Download Citation  Download Citation
  Print
Share:      
METRICS
1
Crossref
0
Scopus
4,233
View
126
Download
Related articles
The Study on the Validation of AMA Pure-Tone Audiometer  2009 December;5(1)
Audiologic Evaluation of Central Auditory Processing Disorders  2010 June;6(1)
Audiometric Calibration of Speech Audiometer  2016 July;12(3)
Editorial Office
RN. 8602, Hallym University,
1 Hallymdaehak-gil, Chuncheon-si, Gangwon-do 24252, Korea
TEL: +82-70-8274-4268   FAX: +82-33-256-3420   E-mail: audiologykorea@gmail.com
About |  Browse Articles |  Current Issue |  For Authors
Copyright © Korean Academy of Audiology.                 Developed in M2PI